# Rigol scope stats in Go
|
||||||
|
|
||||||
|
This tool allows you to pull measurement statistics from a Rigol Oscilloscope over the network connection and outputs them in CSV format. I've tested it with my DS1054Z so far.
|
||||||
|
|
||||||
|
It can also take screenshots at the same intervals.
|
||||||
|
|
||||||
|
## Usage
|
||||||
|
|
||||||
|
# go run .\rigol.go -host 192.168.7.90 -channels 2 -clear -interval 60 -vmax -vmin -screen > file.csv
|
||||||
|
|
||||||
|
Options are as follows. If you have more than 5 statistics displayed (on my scope at least) then reading them takes a few seconds as it has to recalculate the values as the others scroll off the list.
|
||||||
|
|
||||||
|
-channels int
|
||||||
|
number of channels to collect (default 4)
|
||||||
|
-clear
|
||||||
|
clear stats after collection
|
||||||
|
-count int
|
||||||
|
number of measurements to take. -1 = no limit (default -1)
|
||||||
|
-freq
|
||||||
|
include frequency
|
||||||
|
-host string
|
||||||
|
hostname or IP address (default "rigol")
|
||||||
|
-interval int
|
||||||
|
number of seconds between readings (default 1)
|
||||||
|
-port int
|
||||||
|
tcp port to use (default 5555)
|
||||||
|
-screen
|
||||||
|
collect screenshots in PNG format
|
||||||
|
-vavg
|
||||||
|
include Vavg (default true)
|
||||||
|
-vmax
|
||||||
|
include Vmax
|
||||||
|
-vmin
|
||||||
|
include Vmin
|
||||||
|
-vpp
|
||||||
|
include Vpp
|
||||||
|
-vrms
|
||||||
|
include Vrms
|
||||||
|
|
||||||
|
## Setup
|
||||||
|
|
||||||
|
You can configure the scope over the network too - this is especially useful if you want to set specific options across a number of tests.
|
||||||
|
|
||||||
|
I use the following example quite frequently - telnet to the scope on port 5555 and send the following commands. It's worth doing a few at a time or sometimes one will get missed if you paste in too many lines at once.
|
||||||
|
|
||||||
|
*RST
|
||||||
|
|
||||||
|
:TIM:MODE ROLL
|
||||||
|
:TIM:MAIN:SCAL 5
|
||||||
|
:ACQ:TYPE HRES
|
||||||
|
:DISP:TYPE DOTS
|
||||||
|
|
||||||
|
:CHAN1:BWL ON
|
||||||
|
:CHAN1:UNIT AMP
|
||||||
|
:CHAN1:COUP DC
|
||||||
|
:CHAN1:PROB 100
|
||||||
|
:CHAN1:RANG 4
|
||||||
|
:CHAN1:OFFS -1.5
|
||||||
|
:CHAN1:DISP ON
|
||||||
|
|
||||||
|
:CHAN2:BWL ON
|
||||||
|
:CHAN2:UNIT VOLT
|
||||||
|
:CHAN2:COUP DC
|
||||||
|
:CHAN2:PROB 1
|
||||||
|
:CHAN2:RANG 8
|
||||||
|
:CHAN2:OFFS -3
|
||||||
|
:CHAN2:DISP ON
|
||||||
|
|
||||||
|
:MEAS:STAT:ITEM VAVG,CHAN1
|
||||||
|
:MEAS:STAT:ITEM VAVG,CHAN2
|
||||||
|
|
||||||
|
:CURS:MODE TRAC
|
||||||
|
:CURS:TRAC:SOURce1 CHAN1
|
||||||
|
:CURS:TRAC:SOURce2 CHAN2
|
||||||
|
:CURSor:TRACk:AX 586
|
||||||
|
:CURSor:TRACk:BX 586
|
||||||
|
|
||||||
|
Detail for these can be found in the Rigol Oscilloscope Programming guides, but in summary this sets the scope to a rolling 60s view, reading a range of -0.5 to 3.5A on Channel 1 and -1 to 7 V on channel 2, and enables the cursor for realtime measurements.
|
Loading…